Dual Flame ionization detector (FID) and injectors are installed
• Wide linear dynamic range
• No makeup gas required due to efficient sweeping of column effluent by hydrogen combustion gas
• Air flow designed to minimize contamination and residue buildup
• 1/8-in. fittings
• PPC pneumatics – software flow control of hydrogen and air
• Flame out warning and ready interlock
• Auto ignite if flame out detected
Operating temperature: 100 ˚C to 450 ˚C in 1 ˚C increments
Sensitivity: > 0.015 coulombs/g C
Minimum detectable quantity: < 3 • 10-12 g C/sec octane at a S/N = 2 to 1
Linearity: > 10^6
Signal filtration: 50, 200, 800 msec
Input range: 1, 20
Makeup gas: Not required
The touch-screen graphical user interface incorporates
a number of key features:
• Multi-language support (i.e. English, French, Italian,
German, Spanish, Japanese, Chinese and Russian)
• Real-time graphic display of chromatogram
• Injection countdown for manual injections
• Column pressure/flow/velocity calculator
• More upgradable firmware
• Preventative maintenance counter
• Password protection
• Graphical display of temperature and pneumatic
programs
• Status-summary screen
• Log file
• Resolution: 240 x 320
• 256-color display
• Meaningful error/alarm messages
Various user manuals are included along with CTC analytics Peltier thermocouple.