Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
Temperature range from 5 °C (9 ºF) above room temperature to 100 °C (212 °F)
DS control with integrated timer 0 to 99 hrs
Digital temperature setting with an accuracy of a tenth of a degree
Independent temperature safety device class 3.1 (DIN 12880) with optical alarm
Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm (1.97 inch)
Inner glass doors
RS 422 interface for communication software APT-COM™ DataControlSystem
2 chrome-plated racks included
BINDER technical specification conformance certificate
BD 240
Exterior dimensions
Width (mm/inch) 1035 / 40.7
Height (incl. feet) (mm/inch) 825 / 32.5
Depth (mm/inch) 745 / 29.3
Plus door handle and connection (mm/inch) 90 / 3.5
Wall clearance, rear (mm/inch) 100 / 3.9
Wall clearance, side (mm/inch) 160 / 6.3
Steam space volume (l/cu.ft.) 283 / 10.0
Number of doors (ea.) 2
Interior dimensions
Width (mm/inch) 800 / 31.5
Height (mm/inch) 600 / 23.6
Depth (mm/inch) 500 / 19.7
Interior volume (l/cu.ft.) 240 / 8.6
Racks (number standard/max.) 2 / 7
Load per rack (kg/lbs.) 30 / 66
Permitted total load (kg/lbs.) 70 / 155
Weight (empty) (kg/lbs.) 93 / 205
Temperature data
Temperature range from 5 °C (9 ºF) above ambient temperature to (°C/°F)
100 / 212
Temperature variation
at 37 °C (98.6 °F) (± K) 0.5
at 50 °C (122 °F) (± K) 0.9
Temperature fluctuation
at 37 °C (98.6 °F) (± K) 0.1
at 50 °C (122 °F) (± K) 0.1
Warm-up time 1)
to 37 °C (98.6 °F) (min.) 70
to 50 °C (122 °F) (min.) 115
Recovery time after doors were open for 30 sec. 1)
at 37 °C (98.6 °F) (min.) 5
at 50 °C (122 °F) (min.) 6
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10%) 50 / 60 Hz (V) 230
Nominal power (kW) 0.68
Energy consumption at 37 °C (98.6 °F) (W) 33